UHPLC-MS vs HPLC: what each one measures

Table of Contents

Sign-up Now

To receive updates and offers
Liquid chromatography mass spectrometry instrument in the BioRegen laboratory

HPLC and UHPLC-MS are both liquid chromatography: they separate a sample into its components. The difference is what happens next. HPLC with a UV detector tells you how much signal sits in each peak. UHPLC-MS also tells you the mass of what is in it, which is what makes an identity result possible.

Key points

  • Both separate a sample on a column. UHPLC uses finer particles at higher pressure, giving sharper peaks and shorter runs.
  • HPLC with UV detection measures absorbance. It cannot tell two compounds apart if they leave the column at the same time.
  • Mass spectrometry measures mass, so identity can be confirmed in the same run as purity.
  • BioRegen reports purity, identity and net content by UHPLC-MS, the name printed on every certificate.

Chromatography in one paragraph

A dissolved sample is pumped through a packed column. Each component interacts with the packing differently, so components leave the column at different times. The time a component takes is its retention time, and the record of signal against time is the chromatogram. Everything else is about how well peaks are separated and how they are detected.

What the “U” changes

Ultra high performance liquid chromatography uses columns packed with smaller particles, typically under 2 µm, at far higher pressure than conventional HPLC. The result is narrower peaks, better separation of components that elute close together, and shorter runs. Closer peaks resolved means a purity figure that is less likely to hide a component under the main peak.

Liquid chromatography mass spectrometry instrument in the BioRegen laboratory
The UHPLC-MS system in the BioRegen laboratory.

What the mass spectrometer adds

After separation, the mass spectrometer ionizes each component and measures its mass to charge ratio. That adds a second, independent dimension to every peak:

  • Identity. The measured mass is compared with the expected mass of the compound named at submission, together with retention time.
  • Co-eluting components. Two compounds leaving the column together show as one UV peak, but as different masses.
  • Wrong compound, right time. A different compound that happens to elute where the expected one would is caught by its mass.

Side by side

HPLC with UVUHPLC-MS
SeparationLarger particles, lower pressureSub 2 µm particles, higher pressure, sharper peaks
DetectionUV absorbanceMass to charge ratio
PurityMain peak share of UV signalMain peak share of all peaks detected
IdentityOnly by matching a reference standard’s retention timeBy measured mass and retention time, in the same run
Same-time componentsCan merge into one peakDistinguished by mass

Why an HPLC purity figure can mislead

A UV chromatogram with one large peak looks convincing. Without a reference standard run alongside, it does not show what that peak is. A sample of the wrong compound, or a mixture that co-elutes, can still produce a single clean peak and a high purity figure. Purity and identity answer different questions, and HPLC with UV alone answers only the first.

Different detectors, different numbers

UV and mass spectrometry respond differently to different components, so a purity figure from one is not always identical to a figure from the other on the same sample. Compare like with like, and look for the method on the certificate.

Where HPLC still fits

HPLC with UV is a long established, robust technique. With a qualified reference standard it supports both identity by retention time and quantitation, and it remains common in quality control. The limits above apply when it is used alone, without a standard or a second detector.

How BioRegen reports

  1. Purity

    Main peak as a percentage of all peaks detected by UHPLC-MS, printed to two decimals.

  2. Identity

    Measured mass and retention time compared with the specification for the compound named at submission.

  3. Net content

    When ordered, milligrams in the vial from a four to five point calibration with reference standards, with R² printed on the certificate.

Purity and identity cost $165 per sample, or $230 with net content. Results take 2 to 5 business days.

Questions

Is UPLC the same as UHPLC?

Yes, in practice. UPLC is one instrument maker’s name for ultra high performance liquid chromatography; UHPLC is the general term, and the one BioRegen uses on its site and certificates.

Is LC-MS the same as UHPLC-MS?

LC-MS is any liquid chromatography coupled to mass spectrometry. UHPLC-MS is LC-MS with ultra high performance separation.

Can mass spectrometry tell every compound apart?

Not always. Compounds with the same mass, such as some isomers, need separation by retention time or another technique. The certificate reports what the method can establish.

Why does the mass spectrum show a peak at about half the molecular weight?

Peptides often carry two or more charges, and the instrument measures mass divided by charge. See how to read a certificate.

For research use only. BioRegen certificates report analytical results for the sample tested and make no claim about the use of any compound.

Share this Article

Related reading

Analyst holding a petri dish in a laboratory, graded in BioRegen green
Endotoxin units, why results start with "<", how dilution sets the reporting level, and EU/mL to EU/mg.
Sample BioRegen certificate for a BPC-157 and TB-500 blend
Identity and purity for every component, composite chromatograms, and the named blends you can pick.
Laboratory bench with a laptop, glassware and microscope
Pooled results from every certificate the lab has issued: purity, identity, net content and endotoxin, by compound, updated automatically.