What EU/mL means on an endotoxin report

Table of Contents

Sign-up Now

To receive updates and offers
Analyst holding a petri dish in a laboratory, graded in BioRegen green

Endotoxin results are reported in EU/mL, and most of them start with a “<” sign. Both are easy to misread. This guide explains what the unit measures, why the lowest reportable level changes with dilution, and how to convert a result to EU per mg of compound.

Key points

  • An endotoxin unit (EU) expresses endotoxin activity against a reference standard. EU/mL is that activity per mL of the solution tested.
  • A result with “<” means endotoxin was not detected. The figure is the lowest level the assay can report at the dilution tested.
  • Every tenfold dilution raises that lowest reportable level tenfold.
  • Dividing EU/mL by the reconstituted concentration in mg/mL gives EU/mg.

What an endotoxin unit measures

Endotoxins are components of the outer membrane of certain bacteria. They are not measured by weight. The LAL assay measures their biological activity and expresses it in endotoxin units, calibrated against a reference standard. EU/mL is the activity found in each mL of the solution placed in the test.

How the lab measures it

BioRegen runs a kinetic chromogenic LAL assay on an Endosafe nexgen-PTS (see endotoxin testing for price and turnaround). Each cartridge reads between 0.05 and 5.0 EU/mL and runs the sample alongside a spiked copy of itself. A run is reported only when it passes its own checks:

  • Spike recovery between 50% and 200%, which shows the sample is not hiding or inflating the signal.
  • Replicate agreement, with a coefficient of variation below 25%.
  • Test suitability confirmed by the instrument.

Why most results start with “<“

When no endotoxin is detected, the assay cannot print zero. It prints the lowest level it could have seen, with a “<” in front. A result of <0.05 EU/mL means not detected, at a level where 0.05 EU/mL would have been seen.

Across 173 endotoxin reports issued by the lab, 99% were below the lowest reportable level of the assay.

Why dilution changes the number

Peptide solutions can interfere with the LAL reaction. Diluting the sample removes that interference, and spike recovery confirms it worked. The cartridge range stays the same, so every tenfold dilution raises the lowest level that can be reported in the original solution tenfold.

DilutionLowest reportablePer mg at 10 mg/mL
1:1, undiluted<0.05 EU/mL<0.005 EU/mg
1:10<0.5 EU/mL<0.05 EU/mg
1:100<5 EU/mL<0.5 EU/mg
1:1,000<50 EU/mL<5 EU/mg
Nominal values for a cartridge reading down to 0.05 EU/mL. The exact level depends on the cartridge and is printed on each report.

A “<” result can be compared only with a limit above it. If you work to a limit of your own, tell the lab when you submit, so the dilution can be chosen to reach it.

From EU/mL to EU/mg

For a powder, the report gives the result for the solution as reconstituted and per mg of compound. The conversion is one division.

Worked example

  1. Vial reconstituted10 mg in 1 mL = 10 mg/mL
  2. Dilution needed for a valid run1:5,000
  3. Lowest reportable at that dilution<302 EU/mL
  4. Per mg: 302 ÷ 10<30.2 EU/mg

Drawn from a real report, with the compound removed. The cartridge used read slightly above 0.05 EU/mL at its low end, hence 302 rather than 250.

What the report does not say

Current BioRegen endotoxin reports state the measured value, the dilution, the calculation and the validity checks. They make no statement of conformity with any limit. How a result compares with a limit is for the reader to judge, using the figures printed.

Endotoxin is its own test

Purity and identity by UHPLC-MS say nothing about endotoxin, and an endotoxin result says nothing about purity or identity. Each needs its own certificate.

Questions

What does <0.05 EU/mL mean?

Endotoxin was not detected in the undiluted solution, at a level where 0.05 EU/mL would have been seen.

Why was my sample diluted?

So the run would pass its validity checks. Undiluted peptide solutions often interfere with the LAL reaction; spike recovery shows when the dilution has removed that interference.

Is a higher “<” value a worse result?

No. It means not detected at a higher reporting level, set by the dilution. It does not mean endotoxin was found.

What does endotoxin testing cost?

$195 per sample. pH is $45 and is issued on its own certificate. Results take 2 to 5 business days.

For research use only. Endotoxin reports state analytical results for the sample tested. They are not specifications and make no claim about the use of any compound.

Share this Article

Related reading

Liquid chromatography mass spectrometry instrument in the BioRegen laboratory
How the two techniques separate and detect, why identity needs mass, and what that means for a purity figure.
Sample BioRegen certificate for a BPC-157 and TB-500 blend
Identity and purity for every component, composite chromatograms, and the named blends you can pick.
Laboratory bench with a laptop, glassware and microscope
Pooled results from every certificate the lab has issued: purity, identity, net content and endotoxin, by compound, updated automatically.